**气体检测管37LL 检测范围0.3-4.5ppm

**气体检测管37LL 检测范围0.3-4.5ppm
Performance
Measuring Range
0.3 to 4.5 ppm
Number of Pump Strokes
4
Correction Factor
1
Sampling Time 2 minutes per pump stroke
Detecting Limit 0.2 ppm (n=4)
Colour Change White → Pale purple
Reaction Principle CHCl3 + I2O5 + H2S2O7 → Cl2
Cl2 + 3.3'-DIM**NAPHTHIDINE ⅹ reaction product
Coefficient of Variation 10% (for 0.3 to 1.5 ppm), 5% (for 1.5 to 4.5 ppm)
Shelf Life 1 Year
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coe**sting substances and their interferences (NOTE)
Substance Conce**ation Interference Change colour by itself to
Chlorine ℃6ppm + Pale purple
Bromine ″4.5ppm No No
Iodine ″4.5ppm No No
1,1,1-trichloroethane   + Pale purple
trichloro**ene   + Pale purple
Calibration gas generation Diffusion tube method
T**-TWA T**-STEL Explosive range
10 ppm - -
仕様
目盛範囲: 0.3~4.5ppm〔基準n=4〕
測定範囲: 0.3~4.5ppm〔基準n=4〕
吸引回数〔n〕: 4〔基準〕
1回(100ml)の吸引時間: 2分
検知限度: 0.2ppm〔n=4〕
変色: 白色→淡紫色
温?湿度補正: 温度
有効期限: 1年(冷蔵庫保存)
指示精度: CV=5~10%
(CV:変動係数=σ:標準偏差÷平均値×100)
干渉ガス
ガス名 共存の場合の濃度と結果 単独の場合
塩素 2倍以上 +誤差 淡紫色に変色
1,1,1-トリクロロエタン   +誤差 淡紫色に変色
トリクロロエチレン   +誤差 淡紫色に変色
校正用ガス ガス拡散管法
許容濃度 管理濃度 TWA STEL 爆発範囲